The characteristics of the URIT H12 hemoglobin test strips, used with the URIT-12 hemoglobin meter, are as follows:
Sample Type: Capillary or venous whole blood (13 to 15 µL, equivalent to one drop of blood).
Measurement Principle: Optical reflectance (dry chemistry colorimetric method, azidomethemoglobin method).
Measured Parameters: Hemoglobin and hematocrit (HCT).
Measurement Speed: Results obtained in 10 seconds or less.
Measurement Range: 4.0 g/dL to 24.0 g/dL (results outside this range are indicated by “Lo” or “Hi”).
Accuracy: ≥ 95%.
Calibration: Factory calibrated according to the ICSH (International Council for Standardization in Haematology) standard.
Memory: Can store up to 250 test results.
Automation: Self-checking, automatic judgment, and malfunction display.
Strip Storage Conditions: Store in a dry, dark place between 0°C and 30°C, away from direct sunlight and heat, and not frozen.
Use: Strips should be used immediately after removal from the container; used strips should be disposed of as normal medical waste.
Restrictions on Use: Not for serum, plasma, arterial blood, or patient self-testing; intended for use by healthcare professionals.